Variations of MCT photoconductors performance with background flux density

机译:MCT光电导体性能随背景通量密度的变化

获取原文
获取原文并翻译 | 示例

摘要

Variations of peak detectivity D*(λ_p) and peak responsivity R_v(λ_p) values with background flux density have been examined on Photoconductive Mercury-Cadmium Telluride / Hg_(1-x)Cd_xTe (PC MCT) infrared radiation detectors. Measurements were performed at operating temperature T_(op) ≈ 78 K. Performance was measured on multi-element PC MCT infrared radiation detectors (N = 3x32 and 128 elements) with responsivity peak wavelength λ_p from 10,5 to 12,0 μm at the integral background flux density varied from (8-10)E+17 to (6-8)E+15 photon·cm~(-2)·sec~(-1). Tested PC MCT infrared radiation detectors demonstrate background limited performance (BLIP).
机译:在光导汞-碲化镉/ Hg_(1-x)Cd_xTe(PC MCT)红外辐射探测器上检查了峰值探测灵敏度D *(λ_p)和峰值响应灵敏度R_v(λ_p)值随背景通量密度的变化。在工作温度T_(op)≈78 K下进行测量。在多元素PC MCT红外辐射检测器(N = 3x32和128个元素)上测量性能,响应峰值波长λ_p在10,5至1,20μm。积分背景通量密度从(8-10)E + 17到(6-8)E + 15光子·cm〜(-2)·sec〜(-1)不等。经过测试的PC MCT红外辐射探测器证明了背景受限的性能(BLIP)。