Radiometric calibration of an airborne multispectral scanner

机译:机载多光谱扫描仪的辐射定标

获取原文
获取原文并翻译 | 示例

摘要

Abstract: The absolute radiometric calibration of the NS001 thematic mapper simulator reflective channels was examined based on laboratory tests and in-flight comparisons to ground measurements. The NS001 data are calibrated in-flight by reference to the NS001 internal integrating sphere source. This source's power supply or monitoring circuitry exhibited greater instability in-flight during 1988-1989 than in the laboratory. Extrapolating laboratory behavior to in-flight data resulted in 7-20% radiance errors relative to ground measurements and atmospheric modeling. Assuming constancy in the source's output between laboratory and in-flight resulted in generally smaller errors. Upgrades to the source's power supply and monitoring circuitry in 1990 improved its in-flight stability, though in-flight ground reflectance based calibration tests have not yet been performed.!
机译:摘要:基于实验室测试以及在飞行中与地面测量值的比较,检查了NS001专题测绘仪模拟器反射通道的绝对辐射校准。参照NS001内部积分球源在飞行中校准NS001数据。 1988-1989年期间,此电源的电源或监视电路在飞行中比在实验室中表现出更大的不稳定性。将实验室行为推断为飞行中的数据,相对于地面测量和大气建模,导致7-20%的辐射误差。假设实验室和飞行中源的输出保持恒定,通常会导致较小的误差。尽管尚未进行基于机内地面反射率的校准测试,但1990年对信号源的电源和监视电路进行了升级,提高了其机内稳定性。