首页> 外文会议>Conference on Biomedical Optoacoustics Ⅲ Jan 20-21, 2002 San Jose, USA >Two-dimensional detection of optoacoustic stress transients
【24h】

Two-dimensional detection of optoacoustic stress transients

机译:二维检测光声应力瞬变

获取原文
获取原文并翻译 | 示例

摘要

For enhanced optoacoustic imaging in biomedical applications more than one-dimensional detection is required. Tomographic images comparable to ultrasound B-scans can be generated with linear arrays. This work presents a detection scheme for such an array based on piezoelectric films. Each single detector consists of a ringshaped active area with a diameter of less than one milimeter, which leads to a high lateral resolution. Because of the small dimensions of the systems it is suitable for applications with limited accessibility like ophthalmic or endoscopic use. The sensitivity of a single detector is close to 0.5 mV/bar. First measurements on layered tissue phantoms made of gelatine and absorbing films show the potential of such an array for depth profiling as well as for two-dimensional imaging of simple structures.
机译:为了在生物医学应用中增强光声成像,不仅仅需要一维检测。可以用线性阵列生成与超声B扫描相当的断层图像。这项工作提出了一种基于压电薄膜的阵列检测方案。每个单个检测器都包括一个直径小于一毫米的环形活动区域,这导致了较高的横向分辨率。由于系统的尺寸小,它适用于可访问性有限的应用,例如眼科或内窥镜。单个检测器的灵敏度接近0.5 mV / bar。对由明胶和吸收膜制成的分层组织模型的首次测量表明,这种阵列可用于深度剖析以及对简单结构进行二维成像。