We describe the mechanical, optomechanical and thermal design and development of the IRIS2, an infrared imagcr and spectrograph for operation at the Cassegrain focus of the Anglo Australian Telescope. IRIS2 is reconfigured by four encoded worm driven wheels which carry slits and slit masks, filters, cold stops, and grisms, and a pupil imager. A detector translator provides fine focus. The instrument is housed in a split, or dual, vacuum vessel. Helium cryo-coolers provide operational cooling, but to reduce turn around time during commissioning and maintenance a liquid nitrogen pre-cooling system has been implemented in the main vessel. The slit wheel is housed in a separate, smaller vessel, which may be thermally cycled when new slit masks are installed, while the rest of the instrument remains at operational temperature. The common plate between the vessels serves as the structural base on which the instrument is assembled. Matched trusses on opposite sides of the plate minimise the relative deflection between the slit wheel assembly and the spectrograph optics.
机译:我们描述了IRIS2的机械,光学机械和热学设计和开发,这是一个用于在盎格鲁澳大利亚望远镜的Cassegrain焦点上操作的红外成像仪和光谱仪。 IRIS2通过四个编码的蜗杆驱动轮进行了重新配置,这些轮带有狭缝和狭缝罩,滤光片,冷挡和磨粒,以及一个瞳孔成像器。检测器转换器可提供良好的聚焦。仪器容纳在分体式或双重真空容器中。氦低温冷却器可提供运行冷却,但为减少调试和维护期间的周转时间,已在主容器中实施了液氮预冷却系统。裂隙轮装在一个单独的较小容器中,安装新的裂隙面罩时,该裂隙轮可进行热循环,而仪器的其余部分保持在工作温度下。容器之间的共用板充当仪器在其上组装的结构基础。平板相对两侧的相配桁架将狭缝轮组件和光谱仪光学器件之间的相对偏斜降至最低。