Redox Potential Measurement in Aqueous Solutions and Biological Media

机译:水溶液和生物介质中的氧化还原电位测量

获取原文
获取原文并翻译 | 示例

摘要

Electrochemical methods of redox potential measurements of aqueous solutions and biological media with a platinum electrode were analyzed. Probable reasons for the irreproducibility of open circuit potential measurements were considered. We suggested that accumulation of mixed oxides on the surface of platinum after each measurement may lead to potential shifts of the platinum electrode and therefore to a systematic error in redox potential measurements. Preliminary electrochemical treatment of platinum electrodes by cyclic impulses of externally applied potential was developed from the above considerations. The proposed method of measurement was tested in aqueous solutions of inorganic salts, blood, plasma and blood serum. Time dependencies of platinum electrode potential in the presence of inorganic salts and biological liquids were analyzed. We suggested defining the redox potentials of the tested solutions as potential of the platinum electrode after 10-30 min.
机译:分析了用铂电极测量水溶液和生物介质氧化还原电位的电化学方法。考虑了开路电势测量不可重复的可能原因。我们建议,每次测量后,混合氧化物在铂表面上的积累可能会导致铂电极的电势漂移,从而导致氧化还原电势测量中的系统误差。基于上述考虑,开发了通过外部施加电势的循环脉冲对铂电极进行的初步电化学处理。在无机盐,血液,血浆和血清的水溶液中测试了建议的测量方法。分析了在无机盐和生物液体存在下铂电极电位的时间依赖性。我们建议在10-30分钟后将测试溶液的氧化还原电势定义为铂电极的电势。