首页> 外文会议>ICO20: Optical Devices and Instruments >The Analysis of Optical-electro Collimated Light Tube Measurement System
【24h】

The Analysis of Optical-electro Collimated Light Tube Measurement System

机译:光电准直灯管测量系统分析

获取原文
获取原文并翻译 | 示例

摘要

A new type of collimated light tube (CLT) is mentioned in this paper. The analysis and structure of CLT are described detail. The reticle and discrimination board are replaced by a optical-electro graphics generator, or DLP-Digital Light Processor. DLP gives all kinds of graphics controlled by computer, the lighting surface lies on the focus of the CLT. The rays of light pass through the CLT, and the tested products, the image of aim is received by variant focus objective CCD camera, the image can be processed by computer, then, some basic optical parameters will be obtained, such as optical aberration, image slope, etc. At the same time, motorized translation stage carry the DLP moving to simulate the limited distance. The grating ruler records the displacement of the DLP. The key technique is optical-electro auto-focus, the best imaging quality can be gotten by moving 6-D motorized positioning stage. Some principal questions can be solved in this device, for example, the aim generating, the structure of receiving system and optical matching.
机译:本文提出了一种新型的准直光管(CLT)。详细介绍了CLT的分析和结构。标线和识别板被光电图形生成器或DLP数字光处理器取代。 DLP提供了由计算机控制的各种图形,照明表面位于CLT的焦点上。光线穿过CLT,然后进入被测产品,通过变焦点物镜CCD相机接收到目标图像,该图像可以通过计算机进行处理,然后获得一些基本的光学参数,例如光学像差,图像倾斜度等。同时,电动平移台会移动DLP以模拟有限的距离。光栅尺记录DLP的位移。关键技术是光电自动对焦,通过移动6D电动定位平台可以获得最佳成像质量。该设备可以解决一些主要问题,例如目标生成,接收系统的结构和光学匹配。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号