首页> 外文会议>Medical Imaging VI: Instrumentation >Volume CT of anthropomorphic phantoms using a radiation therapy simulator
【24h】

Volume CT of anthropomorphic phantoms using a radiation therapy simulator

机译:使用放射治疗模拟器模拟拟人体模

获取原文
获取原文并翻译 | 示例

摘要

Abstract: We interface a Toshiba LX-40A radiation therapy simulator with a custom designed data acquisition system and a 386-PC. The LX-40A includes a 14 inch (35.6 cm) x-ray image intensifier and a 1 inch saticon camera. Two-dimensional cone beam projections of an anthropomorphic chest/lung phantom are collected. The Feldkamp algorithm reconstructs the volume data set into a volume image of the phantom. We discuss several volume computed tomography issues including spatial distortions in the x-ray image intensifier, scatter, and veiling glare. We compare the volume computed tomography results with two-dimensional CT imaging of the same phantom scanned on the equipment. !26
机译:摘要:我们将东芝LX-40A放射治疗模拟器与定制设计的数据采集系统和386-PC连接。 LX-40A包括14英寸(35.6厘米)X射线图像增强器和1英寸saticon相机。收集拟人化胸部/肺部幻影的二维锥束投影。 Feldkamp算法将体积数据集重建为体模的体积图像。我们讨论了几个体积计算机断层扫描问题,包括X射线图像增强器中的空间畸变,散射和遮盖眩光。我们将体积计算机断层扫描结果与在设备上扫描的相同体模的二维CT成像进行比较。 !26