A geometry method for evaluating the values of intraocular lens tilt and decentration

机译:用于评估人工晶状体倾斜和偏斜值的几何方法

获取原文

摘要

PURPOSE: To propose a new method to evaluate the values of intraocular lens (IOL) tilt and decentration with anterior segment optical coherence tomography (AS-OCT). METHODS: Based on postoperative anterior segment images of IOL from two patients, the values of IOL tilt and decentration were evaluated through curve fitting, image registration and basic geometric theories. RESULTS: The angles of IOL tilt were 2.200 degrees and 2.688 degrees respectively, and the length of IOL decentration were 0.555 mm and 0.474 mm, which were all credible, compared with clinical observations. CONCLUSIONS: Our Geometry-based method to evaluate the values of IOL tilt and decentration might be reasonable and simple with a good repeatability, which may be proved available in other circumstances — IOL implanting and evaluating the different effects on various types of IOL implantation.
机译:目的:提出一种新的方法,通过前段光学相干断层扫描(AS-OCT)评估人工晶状体(IOL)的倾斜度和偏心度。方法:根据两名患者的人工晶状体的术后前节图像,通过曲线拟合,图像配准和基本几何理论评估人工晶状体的倾斜度和偏心度值。结果:IOL倾斜角度分别为2.200度和2.688度,IOL偏心长度分别为0.555mm和0.474mm,与临床观察相比均可信。结论:我们基于几何的评估IOL倾斜度和偏心度值的方法可能是合理而简单的,并且具有良好的可重复性,这在其他情况下可能会得到证明-IOL植入和评估不同类型IOL植入的效果。