首页> 外文会议>Junior Conference on Lighting >Calibration of a Brightness Matching Experiment setup in Mesopic and Scotopic Conditions
【24h】

Calibration of a Brightness Matching Experiment setup in Mesopic and Scotopic Conditions

机译:在中视和暗视条件下校准亮度匹配实验装置

获取原文

摘要

The system under calibration has been previously characterized in photopic conditions. Due to increased interests in mesopic conditions, it was adapted to be usable in such an environment. Preliminary tests raised concerns about the stability of the generated stimulus. This paper retraces the steps which were consequently applied to fully capture, analyze and comprehend the source of observed stimulus instabilities. Through statistical analysis of the numerical results, some hypotheses were dismissed. The purpose of this paper is to expose a rigorous methodology in order to establish a well-controlled experimental environment.
机译:先前已在可见光条件下对经过校准的系统进行了表征。由于对介观条件的兴趣增加,因此使其适合在这样的环境中使用。初步测试引起了人们对所产生刺激的稳定性的担忧。本文追溯了后来被用来完全捕获,分析和理解所观察到的刺激不稳定性来源的步骤。通过对数值结果的统计分析,一些假设被驳回了。本文的目的是公开一种严格的方法,以建立一个可控的实验环境。