Microbial growth can be continuously followed up by measuring the interface impedance Z/sub i/ of two electrodes immersed in a culture medium. A new method based on sinusoidal fitting is used to reduce measurement errors. It applies an exact linearization with which it is possible to obtain a linear relationship of the parameters to be estimated. This permits the application of the least squares procedure, producing pure sinusoidal waves, which greatly reduce the error in the determination of the interface impedance amplitude and phase components.
机译:可以通过测量浸入培养基中的两个电极的界面阻抗Z / sub i /来连续跟踪微生物的生长。一种基于正弦拟合的新方法可减少测量误差。它应用了精确的线性化,通过该线性化可以获取要估计的参数的线性关系。这允许应用最小二乘法,从而产生纯正弦波,从而大大减少了确定界面阻抗幅度和相位分量时的误差。