A 13.56MHz wireless power transfer system with a 1X/2X reconfigurable resonant regulating (R3) rectifier and wireless power control for biomedical implants is presented. Output voltage regulation is achieved through two mechanisms: the local PWM loop of the secondary side controls the duty cycle of switching the rectifier between the 1X and 2X modes; and to adapt to load and coupling variations, the duty cycle information encoded in Manchester code is fed back wirelessly to the primary side using a novel backscattering uplink technique to adjust the transmitter power of the primary coil. The primary transmitter and the secondary R3 rectifier are fabricated in 0.35µm CMOS process. The measured maximum received power and receiver efficiency are 102mW and 92.6%, respectively.
机译:提出了一种具有1X / 2X可重构谐振调节(R3)整流器和用于生物医学植入物的无线功率控制的13.56MHz无线功率传输系统。输出电压调节通过两种机制实现:次级侧的本地PWM环路控制在1X和2X模式之间切换整流器的占空比;为了适应负载和耦合变化,使用新颖的反向散射上行链路技术调整曼彻斯特代码中编码的占空比信息,将其无线反馈到初级侧,以调整初级线圈的发射器功率。初级发射器和次级R3整流器均采用0.35µm CMOS工艺制造。测得的最大接收功率和接收器效率分别为102mW和92.6%。