A Cuffless Blood Pressure Monitor for Home Healthcare Systems Monitored by Health Professionals - A cuffless blood pressure monitor

机译:由卫生专业人员监控的家庭医疗保健系统的无齿状血压计 - 一种诱拐血压计

获取原文

摘要

We tested a new model of healthcare that involves monitoring physiological parameters to improve team-based healthcare. The system consisted of unobtrusive monitoring, a database, and interventions by health professionals. This report discusses the core technologies used, focusing on those involved in monitoring. An unobtrusive blood pressure (BP) estimation system, and web-based care system, including its database, were designed and tested. BP was estimated with a cuffless BP monitor combined with electrocardiography and photoplethysmography. The mean difference (MD) of estimated SBP and reference SBP was 0.2 ± 5.8 mmHg and the limits of agreement ranged from -11.2 mmHg to 11.6 mmHg. The MD of estimated DBP and reference DBP was 0.4 ± 5.7 mmHg and the limits of agreement ranged from -10.8 mmHg to 11.6 mmHg. The estimated BP was enough accurate compared with the cuff-based BP. The core technologies operated well. The proposed system may prove effective for home-based healthcare. Further studies are needed to evaluate the entire care system.
机译:我们测试了一个新的医疗保健模式,涉及监测生理参数,以改善基于团队的医疗保健。该系统由卫生专业人员提供不引人注目的监测,数据库和干预措施。本报告讨论了所使用的核心技术,专注于涉及监测的人。设计并测试了一种不引人注心的血压(BP)估计系统和基于网络的护理系统,包括其数据库。用诱齿BP监测器估计BP,与心电图和光学质肌拍相结合。估计的SBP和参考SBP的平均差异(MD)为0.2±5.8mmHg,协议的限制范围为-11.2mmHg至11.6 mmHg。估计的DBP和参考DBP的MD为0.4±5.7mmHg,协议的限制范围为-10.8 mmhg至11.6 mmhg。与基于袖口的BP相比,估计的BP足够准确。核心技术运营良好。拟议的系统可能对家庭的医疗保健证明是有效的。需要进一步研究来评估整个护理系统。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号