We propose a novel method for nonrigid registration of coronary arteries within frames of a fluoroscopic X-ray angiogram sequence with propagated deformation field. The aim is to remove the motion of coronary arteries in order to simplify further registration of the 3D vessel structure obtained from computed tomography angiography, with the x-ray sequence. The Proposed methodology comprises two stages: propagated adjacent pairwise nonrigid registration, and, sequence-wise fixed frame nonrigid registration. In the first stage, a propagated nonrigid transformation reduces the disparity search range for each frame sequentially. In the second stage, nonrigid registration is applied for all frames with a fixed target frame, thus generating a motion-aligned sequence. Experimental evaluation conducted on a set of 7 fluoroscopic angiograms resulted in reduced target registration error, compared to previous methods, showing the effectiveness of the proposed methodology.
机译:我们提出了一种具有繁殖变形场的荧光X射线血管谱序列框架内的冠状动脉内的非抗原登记的新方法。目的是去除冠状动脉的运动,以简化从计算机断层造影血管造影获得的3D容器结构的进一步登记,具有X射线序列。所提出的方法包括两个阶段:传播相邻的一对非重量注册,以及序列 - 方向固定帧非rigid注册。在第一阶段中,传播的非引力变换顺序地减少了每个帧的视差搜索范围。在第二阶段中,对具有固定目标帧的所有帧应用非重物注册,从而生成运动对准序列。与先前的方法相比,在一组7型荧光血管臂上进行的实验评估导致目标登记误差减少,显示了所提出的方法的有效性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号