【24h】

Coherent amplified optical coherence tomography

机译:相干放大光学相干断层扫描

获取原文
获取原文并翻译 | 示例

摘要

A technique to improve the signal-to-noise ratio (SNR) of a high speed 1300 nm swept source optical coherence tomography (SSOCT) system was demonstrated. A semiconductor optical amplifier (SOA) was employed in the sample arm to coherently amplify the weak light back-scattered from sample tissue without increasing laser power illuminated on the sample. The image quality improvement was visualized and quantified by imaging the anterior segment of a rabbit eye at imaging speed of 20,000 A-lines per second. The theory analysis of SNR gain is given followed by the discussion on the technologies that can further improve the SNR gain.
机译:演示了一种提高1300 nm高速扫频光源光学相干断层扫描(SSOCT)系统的信噪比(SNR)的技术。样品臂中采用了半导体光放大器(SOA),以相干地放大从样品组织反向散射的弱光,而不会增加照射在样品上的激光功率。通过以每秒20,000条A线的成像速度对兔眼的前段进行成像,可以可视化和量化图像质量的提高。给出了SNR增益的理论分析,然后讨论了可以进一步提高SNR增益的技术。