In-vitro model for evaluation of pulse oximetry

机译:用于评估脉搏血氧饱和度的体外模型

获取原文
获取原文并翻译 | 示例

摘要

Abstract: An in vitro model with blood circulating in a silicon tubing system and including an artificial arterial bed is an important tool for evaluation of the pulse oximetry technique. The oxygen saturation was measured on an artificial finger using a pulse oximeter (SpO$-2$/) and on blood samples using a hemoximeter (SaO$-2$/). Measurements were performed at different blood flows and at different blood hematocrits. An increase in steady as well as in pulsatile blood flow was followed by an increase in pulse oximeter readings and a better agreement between SpO$-2$/ and SaO$-2$/ readings. After diluting the blood with normal saline (decreased hematocrit) the agreement was further improved. These results indicate that the pulse oximeter signal is related to blood hematocrit and the velocity of blood. The flow-related dependance of SpO$-2$/ was also evaluated in a human model. These results provided evidence that the pulse oximeter signal is dependent on vascular changes.!
机译:摘要:在硅管系统中具有血液循环并具有人造动脉床的体外模型是评估脉搏血氧饱和度技术的重要工具。使用脉搏血氧仪(SpO $ -2 $ /)在人造手指上测量血氧饱和度,使用血氧仪(SaO $ -2 $ /)在血液样本上测量血氧饱和度。在不同的血流量和不同的血细胞比容下进行测量。稳定血流和搏动性血流增加,随后是脉搏血氧饱和度仪读数的增加以及SpO $ -2 $和SaO $ -2 $ /读数之间的更好的一致性。用生理盐水稀释血液(血细胞比容降低)后,协议得到了进一步改善。这些结果表明脉搏血氧仪信号与血细胞比容和血流速度有关。 SpO $ -2 $ /的与流量相关的依赖性也在人体模型中进行了评估。这些结果提供了脉搏血氧饱和度信号取决于血管变化的证据。