首页> 中文期刊> 《计量学报》 >医用电子束测量方法研究进展

医用电子束测量方法研究进展

         

摘要

The state of the vessel of the water calorimetry employed for the measurement of absorbed dose to water for electron beam for radiotherapy is introduced.Electron beams come to a complete stop within a short distance,dose gradients vary greatly from close to zero.A 12 MeV electron beamhas a typical range in water of 3 cm.The shape of the depth dose curve provides a significant constraint to the structure of the water calorimeters vessel,it is of a pancake cylindrical design.The correction factors of the absorbed dose are sensitive to the structure of the vessel,so further investigation is required.Less data avaliable for the comparison between different primary standards.Also a brief introduction for the the water calorimetry vessel at the National Institute of Metrology is given.%介绍了国际上医用电子束水量热计量热芯的发展概况.电子束表面剂量高且分布均匀,剂量建成区窄,在达到最大剂量点后,深度曲线迅速跌落.尤其是小于12 MeV的电子,其最大剂量深度在3 cm以内.电子的剂量学特征使得其量热芯的结构一般为扁平结构.这样的结构使得热传导和微扰修正因子的计算变得复杂.讨论量热芯结构对剂量修正因子的影响以及国际比对的现状.最后介绍了中国计量科学院相关研究的进展.