Self-designed conical point contact plate and clinical application

         

摘要

Objective: To design a new kind of conical point contact plate (CPCP) in treatment of diaphyseal fracture to avoid further dam age to the blood supply of the cortex. Methods: There were 4 pairs of conicles with a diameter of 1. 5-4 mm beneath the plate symmetrically. The nibs of the conicles were embedd ed about 1-1.5 mm in the cortex by the axial force of screws. The conicle provided a gap about 2 mm between the plate and the bone so as not to compress the periosteum and cortex. And the periostem didnt need to be stripped off dur ing operation and there was no need of postoperative external fixation. Results: Forty-two cases with diaphyseal fracture were stabili zed by using CPCP. The patients could do joint exercise without loading after operation. Partial weight bearing could be performed in 4-6 weeks postoperati vely. There was no obvious external callus in these cases. The average time of b one healing was 3 months in the tibia and forearm fracture and 4 months in the f emur. Complications such as infection, non-union or malunion etc. were not foun d. The implant was removed 6-10 months after operation. Conclusions: CPCP has the advantages of drastic reduction of the implant-to-bone interface virtually, elimination of the impairment of the periosteal blood supply, increase of the rate of healing and good stability whi ch meets the need of early functional exercise.