首页> 中文期刊> 《中国物理快报:英文版》 >Pascal Realization by Comb-Spectral-Interferometry Based Refractometer

Pascal Realization by Comb-Spectral-Interferometry Based Refractometer

         

摘要

To break through the limitations of existing pressure standards,which rely on the gravity and toxic mercury,the national metrological institutes prefer a quantum-based pressure standard.Combining the ideal gas law with helium refractivity measurement,we demonstrate a scheme for the realization of the pressure unit.The refractometer is based on a spectral interferometry with an optical frequency comb and a double-spaced vacuum cell.Through fast Fourier transform of the spectral interferograms of the two beams propagating inside and outside the vacuum cell,the helium refractivity can be obtained with a combined standard uncertainty u(n) of 2.9 × 10-9.Moreover,the final u(p) is ~8.7 × 10-6 in a measurement range of several megapascals (MPa).Our apparatus is compact,fast (15 ms for one single measurement) and easy to handle.Furthermore,the measurement uncertainty will be improved to ~1 × 10-9 or lower if a VIPA-based spectrometer is used.The value of u(p) will thus increase to 3 × 10-6 or better in several MPa.

著录项

  • 来源
    《中国物理快报:英文版》 |2018年第10期|85-88|共4页
  • 作者

    Li-Jun Yang; Yan Li;

  • 作者单位

    State Key Laboratory of Precision Measurement Technology & Instruments, Department of Precision Instruments, Tsinghua University, Beijing 100084;

    State Key Laboratory of Precision Measurement Technology & Instruments, Department of Precision Instruments, Tsinghua University, Beijing 100084;

  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号