首页> 中文期刊> 《癌症治疗(英文)》 >Delivered Dose Verification for Lung Cancer Stereotactic Body Radiotherapy Using Cone-Beam CT

Delivered Dose Verification for Lung Cancer Stereotactic Body Radiotherapy Using Cone-Beam CT

         

摘要

Purpose: Verified the delivered dose distribution of lung cancer Stereotactic Body Radiotherapy (SBRT) using the cone-beam CT images. Methods: Twenty lung cancer patients who underwent SBRT with 100 CBCT images were enrolled in this study. Delivered dose distributions were recalculated on CBCT images with the deformed and non-deformed methods, respectively. The planned and delivered dose distributions were compared using the dose-volume histograms. Results: The delivered target coverage (V100) per patient inside target volume deviated on average were 0.83% ± 0.86% and 1.38% ± 1.40% for Pct vs. Pcbct and Pct vs. Pdcbct, respectively. The Conformity Index (CI) and Gradient Index (GI) showed a good agreement among the plans. For the critical organs, only minor differences were observed between the planned dose and the delivered dose. Conclusions: CBCT images were a useful tool for setup and dose delivery verification for lung cancer patients who underwent SBRT.

著录项

  • 来源
    《癌症治疗(英文)》 |2021年第7期|P.453-459|共7页
  • 作者单位

    Department of Radiation Oncology Hunan Academy of Traditional Chinese Medicine Affiliated Hospital Changsha China;

    School of Medical Information and Engineering Guangdong Pharmaceutical University Guangzhou ChinaGuangdong Province Precise Medicine and Big Data Engineering Technology Research Center for Traditional Chinese Medicine Guangzhou China;

    Department of Radiation Oncology Yunnan Tumor Hospital The Third Affiliated Hospital of Kunming Medical University Kunming China;

    The Second Af?liated Hospital of Hainan Medical University Haikou China;

    Department of Radiation Oncology School of Medicine Xiang’an Hospital of Xiamen University Xiamen University Xiamen China;

    Department of Radiation Oncology School of Medicine Xiang’an Hospital of Xiamen University Xiamen University Xiamen China;

    Department of Radiation Oncology School of Medicine Xiang’an Hospital of Xiamen University Xiamen University Xiamen China;

  • 原文格式 PDF
  • 正文语种 chi
  • 中图分类 数学分析;
  • 关键词

    Lung Cancer; Stereotactic Body Radiotherapy; Cone Beam CT; Dose Verification;