首页> 中文期刊> 《食品研究与开发》 >碰撞池电感耦合等离子体质谱法测定农产品中硒含量的方法

碰撞池电感耦合等离子体质谱法测定农产品中硒含量的方法

         

摘要

比较测试了实验室配制的硒溶液和茶叶消解液,发现在测定茶叶中硒含量时,基质效应明显。采用碰撞池技术(CCT)可以消除这些基质效应,本文以铑为内标,在优化用于测定的硒同位素基础上,对CCT气体(7%氢气+93%氦气)流量也进行了优化,得出以78Se为测定同位素,CCT气体流量为5.0 mL/min的条件下,对茶叶、菠菜和大米3种国家标准物质中硒含量的测定的回收率介于91%~98%之间,准确度符合要求,说明该方法可以用于这3类农产品中硒含量的测定。%Compare the Se solution to tea digestion solution, there was obvious matrix effects in the tea diges-tion solution. As known the collided cell technology(CCT) is a commendable method for eliminating this matrix effect. In this article we used Rh as an internal standard, based the optimized isotope of selenium for test, the flow of CCT gas(7%H2+93%He) was optimized either. The results were that under the flow of gas 5.0 mL/min, determination of isotope 78Se as total selenium in the national reference substances of tea, spinach and rice, the recoveries of selenium were between 91%and 98%, the accuracy were met the demands.

著录项

相似文献

  • 中文文献
  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信