摘要

The feedback control system and model of moving mirror scanning system to a tilt compensated rotated Michelson interferometer for Fourier transform infrared (FT-IR) spectrometer are presented in this paper, with a rotary voice coil motor driver. The system performance is analyzed and a PID controller is added to the control system to improve the static and dynamic performance. The experiments have shown that the relative frequency error is 0.67% at wavenumber of 8000 cm-1. This system has shown a good stability and reliability in practical application, and can be used in the FT-IR spectrometer for atmosphere monitoring.%针对一种倾斜补偿摆动式干涉仪,利用旋转式音圈电机作为驱动单元,设计了扫描控制反馈系统,建立了控制系统的数学模型,分析了系统的性能参数,并将PID控制应用于系统的反馈控制设计,提高了系统的稳态性能和动态性能.结果表明,在波数为8000 cm-1时对应干涉图频率的相对误差达到了0.67%,在实际应用中表现出良好的稳定性和可靠性,能够满足大气环境监测对傅立叶变换红外光谱仪的需要.

著录项

  • 来源
    《红外技术》 |2012年第1期|48-52|共5页
  • 作者单位

    中国科学院安徽光学精密机械研究所,环境光学与技术重点实验室,合肥230031;

    中国科学院安徽光学精密机械研究所,环境光学与技术重点实验室,合肥230031;

    中国科学院安徽光学精密机械研究所,环境光学与技术重点实验室,合肥230031;

    中国科学院安徽光学精密机械研究所,环境光学与技术重点实验室,合肥230031;

    中国科学院安徽光学精密机械研究所,环境光学与技术重点实验室,合肥230031;

    中国科学院安徽光学精密机械研究所,环境光学与技术重点实验室,合肥230031;

  • 原文格式 PDF
  • 正文语种 chi
  • 中图分类 红外系统装置;
  • 关键词

    傅立叶变换红外; 干涉仪; 动镜; 扫描系统; 反馈控制; PID;