首页> 中文期刊> 《针灸推拿医学:英文版》 >Study on Burning Temperature Versus Time Graph of Moxa Sticks in a Novel Animal Moxibustion Device

Study on Burning Temperature Versus Time Graph of Moxa Sticks in a Novel Animal Moxibustion Device

         

摘要

Objective: To invent a novel animal experimentation moxa device and intend to solve temperature fluctuations during moxa stick burning in animal moxibustion test. Methods: Smokeless moxa sticks of 4 mm in diameter and 120 mm in length were selected in this trial. The target temperature was(47±1) ℃. The burning temperature versus time graph of moxa sticks during un-interfered time period were recorded when the distances between moxa sticks and thermometer were 2 mm, 3 mm and 4 mm respectively. In addition, the total lengths of burned moxa stick ashes within 0-5 min were recorded. Then the moxa stick burning temperature versus time graph after intervention was measured. Results: The maximum temperatures during test duration when the distances were 2 mm, 3 mm and 4 mm were(47.37±0.25) ℃,(42.62±2.05) ℃ and(35.50±0.40) ℃respectively the total lengths of burned ashes from 0-5 min were(5.1±0.08) mm,(12.17±0.44) mm,(14.8±0.31) mm,(17.25±0.17) mm and(19.82±0.67) mm respectively with a distance of 2 mm between a moxa stick and the testing area, remove 5 mm ashes every 2 min, then the maximum and minimum temperatures within 10 min were(48.37±0.47) ℃and(47.12±0.25) ℃, showing no statistical significance(P=0.12). Conclusion: The novel animal experimentation moxa device can help to maintain relatively constant moxa stick burning temperature during test period.