摘要

The method of isotope dilution thermal ionization mass spectrometry able to trace to SI was developed to accurately measure trace amount of cadmium in the rice powder sample from CCQM-K24 international key comparison organized by CCQM(Comité Consultatif pour la Quantité de Matière, Paris). By comparison with our previous work, there was much improvement in aspects of digestion of the rice powder sample, separation of cadmium from the sample and dry-mass correction. As a result, the blank of this procedure was reduced and the measurement precisions of isotopic ratios of cadmium were increased. In addition, the uncertainty evaluation of the entire process of measurement was profoundly studied. The analytical result of this work (14.53±0.15) nmol/g was in agreement with the certified value.