首页> 中文期刊> 《上海交通大学学报:英文版》 >A NEW ESTIMATION METHOD OF SPATIAL PEAK TEMPORAL AVERAGE INTENSITY AND SPATIAL PEAK PULSE AVERAGE INTENSITY IN HIFU FIELD

A NEW ESTIMATION METHOD OF SPATIAL PEAK TEMPORAL AVERAGE INTENSITY AND SPATIAL PEAK PULSE AVERAGE INTENSITY IN HIFU FIELD

         

摘要

A new estimation method of maximum intensity, such as I spta in CW and I sppa in PW, in high intensity focused ultrasound (HIFU) field for surgery was developed based on the measurements of acousic power and radical distribution of the relative acoustic pressure in the focal plane of the field. The acoustic power is measured by the radiation force balance or by the calorimetry. Using a miniature hydrophone to scan along several radial directions with the same angle intervals through the acoustic axis in the focal plane, several radial relative pressure distribution curves can be measured and thus the special cross section area related to the maximum intensity and be calculated. The quotient of the measured acousitc power to the special cross section area is the maximum acoustic intensity in the field. finally as an example, the acoustic field distribution of a CW system was measured and the maximum acoustic intensity estimated was given.

著录项

相似文献

  • 中文文献
  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号