Comparisons of dosimetric properties between GD-300 series of radiophotoluminescent glass detector: and GR-200 series of thermoluminescent detectors

机译:GD-300系列放射性光致发光玻璃探测器与GR-200系列热致发光探测器的剂量特性比较

获取原文
获取原文并翻译 | 示例
       

摘要

In order to compare the dosimetric properties of two kinds of solid state detectors used for monitoring external exposures,experiments Were performed for the GD-300 series of radiophotoluminescent glass detectors (RPLGD) and the GR-200 series of thermoluminescent detectors (TLD).X-rays and 137 Cs and 60 Co γ-rays were used to irradiate the RPLGDs and TLDs,their dose linearky,uniformity/reproducibility of signal,fading effect and energy dependence were compared.Both kinds of the detectors have generally good dosimetric properties.The dose linearity and energy dependence of the GD-351 dosimeters and the TLDs are nearly the same,but the RPLGD is much better than the TLD in terms of the uniformity and fading effect.