首页> 外文期刊>核技术:英文版 >Monte carlo study on 6 MV photon beams of a CyberKnife stereotactic radiosurgery system
【24h】

Monte carlo study on 6 MV photon beams of a CyberKnife stereotactic radiosurgery system

机译:蒙特卡洛对射波刀立体定向放射外科系统的6 MV光子束的研究

获取原文
获取原文并翻译 | 示例
       

摘要

In this paper,the beam quality and percent depth dose curves for different field sizes of CyberKnife? system were investigated by Monte Carlo simulations using the PENELOPE code,which has been used to simulate 6 MV photon beam.In water phantom,the absolute doses were calculated for Φ10–60 mm collimators,and percent depth dose curves were evaluated for Φ30–60 mm collimators.The agreement of dose distributions of the calculation with measurement was within 3.0%.The mean energy of photon spectrum was 1.46 MeV,and the beam quality index was 0.632,which was slightly smaller than that of measurement.