首页> 中文期刊> 《等离子体科学和技术:英文版》 >High-speed VUV spectroscopy for edge impurity line emission measurements in HL-2A tokamak

High-speed VUV spectroscopy for edge impurity line emission measurements in HL-2A tokamak

         

摘要

A 20 cm focal length normal incidence vacuum ultraviolet(VUV_20 cm) monochromator with a fast time response has been developed for measuring edge impurity line emission in the wavelength range of 300–2000 ? on an HL-2A tokamak. An aberration corrected concave holographic grating with 1200 grooves/mm is adopted in the monochromator, which provides a wavelength dispersion of 40 ? mm^(-1). The aperture is f/4.5. A channel electron multiplier is used as a detector. The time resolution of the system is 17 μs. Wavelength calibration of the system has been done by using a hollow cathode light source in the laboratory with helium and argon gases. The obtained signals of helium and argon spectra are very strong since the inner surface of the monochromator vacuum chamber is blackened and the stray light level is then significantly reduced. The optical property of the system has been examined by scanning the width of the entrance and exit slits. The system is then installed at the mid-port of the HL-2A tokamak and typical line emissions from the HL-2A plasma are measured. Time behaviors of edge impurity line emissions are observed with the fast time response system in different plasma confinement regimes, especially in the H-mode discharges. The result shows that the VUV_20 cm system works very well to measure the edge impurity line emissions in the edge localized modes phase of H-mode discharges.