首页> 中文期刊> 《光谱学与光谱分析》 >痕量挥发性有机物的高场不对称波形离子迁移谱研究

痕量挥发性有机物的高场不对称波形离子迁移谱研究

         

摘要

FAIMS is a fast and high sensitive technique for detecting trace volatile organic compounds. Spectra of acetone, benzene and toluene were obtained on a homemade high-field asymmetric waveform ion mobility spectrometer and they can be easily separated in the spectra. Three xylene isomeric compounds were also successfully separated in FAIMS. Effect of carrier gas flow rate on the ion intensity was analyzed, and the optimal flow rate of carrier gas was 220 L · h-1 which can be used for the optimization of FAIMS instrument. The detection limit for acetone is 100 ng · L-1 that is an order of magnitude lower than the foreign traditional IMS.%高场不对称波形离子迁移谱(high-field asymmetric waveform ion mobility spectrometry,FAIMS)技术是一种快速、高灵敏的痕量物质检测技术.利用自制的高场不对称波形离子迁移谱仪分别对丙酮、苯和甲苯样品进行了检测,结果表明三种挥发性物质在谱图中的分离效果优于传统离子迁移谱仪;另外分别对二甲苯三种同分异构体进行了检测,成功地实现了三者的分离.实验分析了载气流速对离子信号强度的影响,发现流速为220 L·h-1时信号强度最大,此结果可用于仪器的优化设计.获得丙酮的检出限为100ng·L-1,比国外传统离子迁移谱仪低了一个数量级.

著录项

  • 来源
    《光谱学与光谱分析》 |2011年第1期|12-15|共4页
  • 作者单位

    中国科学院合肥智能机械研究所,传感技术国家重点实验室,安徽合肥,230031;

    中国科学技术大学自动化系,安徽合肥,230027;

    中国科学院合肥智能机械研究所,传感技术国家重点实验室,安徽合肥,230031;

    中国科学技术大学自动化系,安徽合肥,230027;

    中国科学院合肥智能机械研究所,传感技术国家重点实验室,安徽合肥,230031;

    中国科学院合肥智能机械研究所,传感技术国家重点实验室,安徽合肥,230031;

    中国科学院合肥智能机械研究所,传感技术国家重点实验室,安徽合肥,230031;

    中国科学技术大学自动化系,安徽合肥,230027;

    中国科学院合肥智能机械研究所,传感技术国家重点实验室,安徽合肥,230031;

    中国科学院合肥智能机械研究所,传感技术国家重点实验室,安徽合肥,230031;

  • 原文格式 PDF
  • 正文语种 chi
  • 中图分类 光化学分析法(光谱分析法);
  • 关键词

    高场不对称波形离子迁移谱; 痕量挥发性有机物;