Impact of Digital Panoramic Radiograph Magnification on Vertical Measurement Accuracy

机译:数字全景射线照相倍率对垂直测量精度的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Objectives. The purpose of this panoramic radiography study was to assess the impact of image magnification on the accuracy of vertical measurements in the posterior mandible. Methods. Six dental implants, inserted in the posterior segments of a resin model, were used as reference objects. Two observers performed implant length measurements using a proprietary viewer with two preset image magnifications: the low (1.9 : 1) and the medium (3.4 : 1) image magnifications. They also measured the implant lengths in two Digital Imaging Communications in Medicine viewers set at low (1.9 : 1), medium (3.4 : 1), and high (10 : 1) image magnifications. Results. The error between the measured length and the real implant length was close to zero for all three viewers and image magnifications. The percentage of measurements equal to the real implant length was the highest (83.3%) for the high image magnification and below 30% for all viewers with the low image magnification. Conclusions. The high and medium image magnifications used in this study allowed accurate vertical measurements, with all three imaging programs, in the posterior segments of a mandibular model. This study suggests that a low image magnification should not be used for vertical measurements on digital panoramic radiographs when planning an implant in the posterior mandible.
机译:目标。这项全景射线照相研究的目的是评估图像放大率对后下颌骨垂直测量准确性的影响。方法。在树脂模型的后段插入了六个牙科植入物,用作参考对象。两名观察员使用专有的观察器对植入物进行了长度测量,该观察器具有两个预设的图像放大倍率:低(1.9:1)和中等(3.4:1)图像放大率。他们还分别在低(1.9:1),中(3.4:1)和高(10:1)的图像放大倍率下测量了两个“医学数字成像通讯”中的植入物长度。结果。对于所有三个观察者和图像放大率,测量的长度与实际植入物长度之间的误差接近于零。对于高图像放大倍率,等于实际植入物长度的测量百分比最高(83.3%),而对于所有低图像放大倍率的观察者,其测量百分比低于30%。结论。在这项研究中使用的高中图像放大倍数允许在下颌模型的后段使用所有三个成像程序进行精确的垂直测量。这项研究表明,在计划下颌骨后植入物时,不应将低图像放大率用于数字全景X射线照片的垂直测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号