摘要

Endovascular aortic aneurysm repair with the EndurantTM stent-graft system has been shown to be safe and effective in high-risk surgical patients with complex suprarenal and/or infrarenal abdominal aortic aneurysm anatomy. The wireformed M-shaped stent architecture and proximal springs with anchoring pins theoretically permit optimal sealing in shorter and more angulated proximal aneurysm necks even under off-label conditions. Nonetheless, extremely difficult anatomical situations and inherent graft system-related limitations must be anticipated. Herein, we describe our techniques to overcome the capture of the tip sleeve within the suprarenal bare-stent anchoring pins, other endograft segments, and native vessels.
机译:已经证明,使用Endurant TM 支架系统修复血管内主动脉瘤对于高风险的复杂肾上和/或肾下腹主动脉瘤解剖结构的外科手术患者是安全有效的。理论上,线形M形支架结构和带有锚定销的近端弹簧即使在没有标签的情况下,也可以在较短且角度更大的近端动脉瘤颈部中实现最佳密封。但是,必须预见极其困难的解剖情况和固有的移植系统相关限制。在本文中,我们描述了克服在肾上半裸支架锚钉,其他内移植段和天然血管中捕获尖端套的技术。