Experimental and numerical investigation into the influence of loading conditions in biomechanical testing of locking plate fracture fixation devices

机译:加载条件对锁定板骨折固定装置生物力学测试影响的实验和数值研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

ObjectivesSecondary fracture healing is strongly influenced by the stiffness of the bone-fixator system. Biomechanical tests are extensively used to investigate stiffness and strength of fixation devices. The stiffness values reported in the literature for locked plating, however, vary by three orders of magnitude. The aim of this study was to examine the influence that the method of restraint and load application has on the stiffness produced, the strain distribution within the bone, and the stresses in the implant for locking plate constructs.
机译:目的次生骨折的愈合受到骨固定器系统刚度的强烈影响。生物力学测试被广泛用于研究固定装置的刚度和强度。然而,文献中报道的用于锁定镀层的刚度值变化了三个数量级。这项研究的目的是研究施加约束和施加载荷的方法对所产生的刚度,骨内的应变分布以及用于锁定板结构的植入物的应力的影响。