Reusable urine glucose sensor based on functionalized graphene oxide conjugated Au electrode with protective layers

机译:基于带保护层的功能化氧化石墨烯共轭金电极的可重复使用的尿葡萄糖传感器

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

An electrochemical based system with multiple layers coated on a functionalized graphene oxide Au electrode was developed to measure glucose concentration in urine in a more stable way. Two types of gold printed circuit boards were fabricated and graphene oxide was immobilized on their surface by chemical adsorption. Multiple layers, composed of a couple of polymers, were uniformly coated on the surface electrode. This device exhibited higher electrochemical responses against glucose, a greater resistivity in the presence of interferential substances in urine, and durable stabilities for longer periods of time than conventional units. The efficiency in current level according to the order and ratio of solution was evaluated during the immobilization of the layer. The fabricated electrodes were then also evaluated using hyperglycemic clinical samples and compared with the patterns of blood glucose measured with commercially available glucose meters. Our findings show that not only was their pattern similar but this similarity is well correlated.
机译:开发了一种基于电化学的系统,该系统在功能化的氧化石墨烯Au电极上涂覆了多层涂层,可以更稳定地测量尿液中的葡萄糖浓度。制备了两种类型的金印刷电路板,并且通过化学吸附将氧化石墨烯固定在其表面上。由一对聚合物组成的多层均匀地涂覆在表面电极上。与常规装置相比,该装置对葡萄糖表现出更高的电化学响应,在尿液中存在干扰物质时具有更高的电阻率,并且在更长的时间内具有持久的稳定性。在固定层的过程中,根据溶液的顺序和比例评估电流水平的效率。然后,还使用高血糖临床样品对制成的电极进行评估,并与使用市售血糖仪测量的血糖模式进行比较。我们的发现表明,不仅它们的模式相似,而且这种相似性也很好地相关。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号