Preclinical Evaluation of a Noncontact Simultaneous Monitoring Method for Respiration and Carotid Pulsation Using Impulse-Radio Ultra-Wideband Radar

机译:基于脉冲无线电超宽带雷达的呼吸和颈动脉搏动非接触式同时监测方法的临床前评估

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

There has been the possibility for respiration and carotid pulsation to be simultaneously monitored from a distance using impulse-radio ultra-wideband (IR-UWB) radar. Therefore, we investigated the validity of simultaneous respiratory rates (RR), pulse rates (PR) and R-R interval measurement using IR-UWB radar. We included 19 patients with a normal sinus rhythm (NSR) and 14 patients with persistent atrial fibrillation (PeAF). The RR, PR, R-R interval and rhythm were obtained simultaneously from the right carotid artery area in a supine position and under normal breathing conditions using IR-UWB radar. There was excellent agreement between the RR obtained by IR-UWB radar and that manually counted by a physician (intraclass correlation coefficient [ICC] 0.852). In the NSR group, there was excellent agreement between the PR (ICC 0.985), average R-R interval (ICC 0.999), and individual R-R interval (ICC 0.910) measured by IR-UWB radar and electrocardiography. In the PeAF group, PR (ICC 0.930), average R-R interval (ICC 0.957) and individual R-R interval (ICC 0.701) also agreed well between the two methods. These results demonstrate that IR-UWB radar can simultaneously monitor respiration, carotid pulse and heart rhythm with high precision and may thus be utilized as a noncontact continuous vital sign monitoring in clinical practice.
机译:使用脉冲无线电超宽带(IR-UWB)雷达可以从远处同时监视呼吸和颈动脉搏动。因此,我们调查了使用IR-UWB雷达同时进行呼吸频率(RR),脉搏频率(PR)和R-R间隔测量的有效性。我们纳入了具有正常窦性心律(NSR)的19例患者和持续性房颤(PeAF)的14例患者。 RR,PR,R-R间隔和心律是使用IR-UWB雷达同时从仰卧位和正常呼吸条件下的右颈动脉区域获得的。 IR-UWB雷达获得的RR与医生手动计算的RR之间具有极好的一致性(类内相关系数[ICC] 0.852)。在NSR组中,通过IR-UWB雷达和心电图测量的PR(ICC 0.985),平均R-R间隔(ICC 0.999)和单个R-R间隔(ICC 0.910)之间具有极好的一致性。在PeAF组中,两种方法之间的PR(ICC 0.930),平均R-R间隔(ICC 0.957)和单个R-R间隔(ICC 0.701)也很吻合。这些结果表明,IR-UWB雷达可以高精度地同时监测呼吸,颈动脉搏动和心律,因此可以在临床实践中用作非接触式连续生命体征监测。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号