Continuous observation of aquatic pCO2  at the ocean surface, with a sensitive response time and high spatiotemporal resolution, is essential for research into the carbon biogeochemical cycle. In this work, a portable tunable diode laser absorption spectroscopy (TDLAS) system for dissolved CO2 detection in surface seawater, coupled with a home-made headspace equilibrator, allowing real time underway measurements, is described. Both the optical detection part and sample extraction part were integrated together into a compact chamber. An empirical equation suitable for this system was acquired, which can convert the concentration from the gas-phase to the aqueous-phase. A monitoring precision of 0.5% was obtained with time-series measurement, and the detection limits of 2.3 ppmv and 0.1 ppmv were determined with 1 s and 128 s averaging time, respectively. Sampling device used in this work was ameliorated so that the response time of system reduced by about 50% compared to the traditional ‘shower head’ system. The fast response time reached the order of 41 s when the final concentration span was 3079 ppmv. For1902 ppmv, this figure was as short as 20 s. Finally, a field underway measurement campaign was carried out and the results were briefly analyzed. Our work proved the feasibility of the TDLAS system for dissolved CO2 rapid detection.
机译:在海洋表面的持续观察,具有敏感的响应时间和高时的分辨率,对碳生物地球化学循环的研究至关重要。在这项工作中,用于表面海水中的用于溶解CO2检测的便携式可调二极管激光吸收光谱(TDLAS)系统,与自制的顶部空间平衡器相结合,允许实时进行测量。光学检测部分和样品提取部分都集成在一起到紧凑的室中。获得适用于该系统的经验方程,其可以将浓度从气相转化为水相。用时间序列测量获得0.5%的监测精度,分别用1S和128秒的平均时间测定2.3ppmV和0.1ppmv的检测限。与传统的“淋浴头”系统相比,在这项工作中使用的采样装置进行了改进,使系统的响应时间减少了大约50%。当最终浓度跨度为3079ppmV时,快速响应时间达到41秒的顺序。对于1902 ppmv,该数字短至20秒。最后,进行了一个现场进行测量运动,并简要分析了结果。我们的工作证明了TDLAS系统的可行性溶解二氧化碳快速检测。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号