Ex-vivo release of Pipeline Embolization Device polytetrafluoroethylene (PTFE) sleeves for improved distal landing zone accuracy in-vivo: A technical note

机译:体内释放管道栓塞装置聚四氟乙烯(PTFE)套管以提高体内远端着陆区的准确性:技术说明

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

BackgroundDistal landing zone accuracy is critical in some intracranial aneurysms treated with the Pipeline Embolization Device (PED), and delayed opening of the distal end of the device can complicate the procedure. Here, we report a technical nuance that facilitates accurate placement of the distal end of the PED by ex-vivo, pre-implantation release of the PED Flex polytetrafluoroethylene (PTFE) sleeves.
机译:背景远端着陆区的准确性在使用管道栓塞装置(PED)处理的某些颅内动脉瘤中至关重要,并且装置远端的延迟打开会使操作复杂化。在这里,我们报告了一项技术上的细微差别,可通过PED Flex聚四氟乙烯(PTFE)套管的离体,植入前释放来促进PED远端的准确放置。