首页> 外文期刊>Applied Spectroscopy >Time-Resolved Step-Scan Infrared Imaging System Utilizing a Linear Array Detector
【24h】

Time-Resolved Step-Scan Infrared Imaging System Utilizing a Linear Array Detector

机译:利用线性阵列检测器的时间分辨步进扫描红外成像系统

获取原文
获取原文并翻译 | 示例
           

摘要

A time-resolved infrared (IR) imaging system combined with a multichannel IR microscope, which utilizes a 16 channel linear array (LA) detector, and step-scan Fourier transform infrared (FT-IR) microscope was developed. The LA detector integrates a readout circuit on each detector element, so the detected signals can be read simultaneously. Thus, this system can perform high speed imaging using the step-scan method, similar to a single channel detector. To verify the capabilities of this system, a reflective sample was examined whose position was altered using a piezo actuator activated by an alternating voltage. In addition, the localization of relaxation dynamics for the liquid crystal (LC) molecules in an LC cell under oscillating electric field conditions was detected by this system.
机译:开发了一种结合了多通道红外显微镜的时间分辨红外成像系统,该系统利用了16通道线性阵列(LA)检测器,并采用了步进扫描傅立叶变换红外(FT-IR)显微镜。 LA检测器在每个检测器元件上集成了一个读出电路,因此可以同时读取检测到的信号。因此,该系统可以使用步进扫描方法执行高速成像,类似于单通道检测器。为了验证该系统的功能,检查了反射样品,该样品的位置使用由交流电压激活的压电致动器进行了更改。另外,通过该系统检测了在振荡电场条件下LC单元中液晶(LC)分子的弛豫动力学的局部。