Computer-assisted antetorsion control prevents malrotation in femoral nailing: an experimental study and preliminary clinical case series

机译:计算机辅助前倾控制可防止股骨钉畸形:一项实验研究和初步临床病例系列

获取原文
获取原文并翻译 | 示例
           

摘要

Materials and methods The effect of fluoroscopy-based navigation for femoral fracture reduction on the prevention of malrotation was examined in an experimental setting followed by a first case series. Eleven cadaver femurs were used. All femurs were reduced by closed methods. An optoelectronic navigation system was utilized to check for fragment reduction and alignment. Fluoroscopic control without navigation was used as the control group. The Six Sigma Analysis [offset capability index (C pk) = 1.3] was used to compare the probability of outliers of more than 15°. In the clinical case series the same navigation tool was used in ten non-consecutive patients with femoral fractures. Torsional differences between both legs were measured postoperatively by CT scan.
机译:材料和方法在随后的第一个病例系列的实验环境中,研究了基于荧光透视的导航对减少股骨骨折的预防旋转不良的影响。使用11具尸体股骨。所有股骨均通过封闭方法复位。利用光电导航系统检查碎片的减少和对齐。对照组采用无导航的荧光镜控制。六西格玛分析[偏移能力指数(C pk )= 1.3]用于比较异常值大于15°的概率。在临床病例系列中,十名非连续性股骨骨折患者使用了相同的导航工具。术后通过CT扫描测量双腿之间的扭转差异。