Dose uniformity of budesonide Easyhaler? under simulated real-life conditions and with low inspiration flow rates

机译:布地奈德Easyhaler的剂量均匀性?在模拟现实条件下且吸气流速低

获取原文
           

摘要

Budesonide Easyhaler? multidose dry powder inhaler is approved for the treatment of asthma. Objectives were to determine the delivered dose (DD) uniformity of budesonide Easyhaler? in simulated real-world conditions and with different inspiration flow rates (IFRs). Three dose delivery studies were performed using 100, 200, and 400 μg/dose strengths of budesonide. Dose uniformity was assessed during in-use periods of 4–6 months after exposure to high temperature (30°C) and humidity (60% relative humidity) and after dropping and vibration testing. The influence of various IFRs (31, 43, and 54 L/min) on the DD was also investigated. Acceptable dose uniformity was declared when mean DD were within 80–120% of expected dose; all data reported descriptively. DD was constant (range: 93–109% of expected dose) at all in-use periods and after exposure to high temperature and humidity for a duration of up to 6 months. DD post-dropping and -vibration were unaffected (range 98–105% of expected dose). Similarly, DD was constant and within 10% of expected dose across all IFRs. Results indicate that budesonide Easyhaler? delivers consistently accurate doses in various real-life conditions. Budesonide Easyhaler? can be expected to consistently deliver a uniform dose and improve asthma control regardless of high temperature and humidity or varying IFR.
机译:布地奈德Easyhaler?多剂量干粉吸入器被批准用于治疗哮喘。目的是确定布地奈德Easyhaler的给药剂量(DD)均匀性?在模拟的现实条件下,并且具有不同的吸气流速(IFR)。使用布地奈德的100、200和400μg/剂量强度进行了三个剂量输送研究。在暴露于高温(30°C)和湿度(60%相对湿度)以及跌落和振动测试之后的4-6个月的使用期间,评估剂量的均匀性。还研究了各种IFR(31、43和54 L / min)对DD的影响。当平均DD在预期剂量的80-120%之内时,宣布剂量可接受。所有数据均以描述性方式报告。在所有使用期间以及暴露于高温高湿下长达6个月之后,DD均保持恒定(范围:预期剂量的93–109%)。滴下后DD和振动不受影响(预期剂量的98–105%)。同样,DD是恒定的,并且在所有IFR中均处于预期剂量的10%以内。结果表明布地奈德Easyhaler?在各种现实生活条件下都能提供始终如一的精确剂量。布地奈德Easyhaler?不论高温高湿或IFR变化如何,都有望持续提供均匀的剂量并改善哮喘控制。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号