Calibration of vector network analyzers on the basis of the LRR-method

机译:基于LRR方法的矢量网络分析仪校准

获取原文
           

摘要

The LRR method for the calibration of vector networkanalyzers is presented. This method belongs to the selfcalibration procedures where the calibration circuits might bepartly unknown. The LRR calibration circuits are all of equalmechanical length in contrast to the well known TRL calibration,which needs a line-standard with a different lengththan the other calibration standards. For the LRR method itis thus not necessary to displace the connectors of the vectornetwork analyzer during calibration in order to contact thecalibration structures. The calibration circuits mainly consistof reflective networks that have to be placed at three consecutivepositions. As the algorithm accounts for different distancesbetween the reflective networks, the circuits are easyto realize. The robust functionality of the LRR method isconfirmed by measurements.
机译:介绍了用于矢量网络分析仪校准的LRR方法。该方法属于自校准程序,其中校准电路可能部分未知。与众所周知的TRL校准相比,LRR校准电路的机械长度均相等,TRL校准需要线标准的长度与其他校准标准的长度不同。因此,对于LRR方法,不需要在校准期间移动矢量网络分析仪的连接器以接触校准结构。校准电路主要由反射网络组成,必须将其放置在三个连续的位置。由于该算法考虑了反射网络之间的不同距离,因此电路易于实现。 LRR方法的强大功能通过测量得到证实。