首页> 外文期刊>Artificial cells, blood substitutes, and immobilization biotechnology. >Alkaline Phosphatase Based Amperometric Biosensor Immobilized by Cysteamine-Glutaraldehyde Modified Self-Assembled Monolayer
【24h】

Alkaline Phosphatase Based Amperometric Biosensor Immobilized by Cysteamine-Glutaraldehyde Modified Self-Assembled Monolayer

机译:半胱胺-戊二醛修饰的自组装单分子膜固定化基于碱性磷酸酶的电流生物传感器

获取原文
           

摘要

: Alkaline phosphatase (ALP) was immobilized with cross-linking agents glutaraldehyde and cysteamine by forming a self-assembled monolayer on a screen printed gold electrode. ALP converts p-nitrophenyl phosphate to p-nitrophenol and phosphate. p-Nitrophenol loses H+ ion and turns into the negatively charged compound p-nitrophenolate at medium pH. As a result, the unstable product formed is measured chronoamperometrically at an application potential of + 0.95 V. The biosensor response depends linearly on p-nitrophenyl phosphate concentration between 0.05 – 0.6 mM with a response time of 40 seconds. Detection limit of the biosensor is 0.033 mM.
机译::通过在丝网印刷金电极上形成自组装单层,将碱性磷酸酶(ALP)固定在交联剂戊二醛和半胱胺上。 ALP将对硝基苯基磷酸酯转化为对硝基苯酚和磷酸酯。对硝基苯酚会损失H +离子,并在中等pH值下变成带负电荷的化合物对硝基苯酚盐。结果,在+0.95 V的施加电势下,通过计时安培法测量了形成的不稳定产物。生物传感器的响应线性依赖于对硝基苯基磷酸酯浓度在0.05 – 0.6 mM之间,响应时间为40秒。生物传感器的检测极限为0.033 mM。