Reduction of Collimator Correction Artefacts with Bayesian Reconstruction in Spect

机译:贝叶斯重构在斑点中减少准直器校正伪像

获取原文
           

摘要

Poor resolution of single photon emission computed tomography (SPECT) has degraded its use in clinical practice. Collimator correction has been shown to improve the reconstructed resolution, but the correction can generate ringing artefacts, which lower image quality. This paper investigates whether Bayesian reconstruction methods could reduce these artefacts. We have applied and tested three Bayesian reconstruction methods: smoothing prior, median root prior, and anatomical prior. To demonstrate the efficacy of these methods, we compared their physical and visual performance both in phantom and patient studies. All the three Bayesian reconstruction methods reduced the collimator correction artefacts. Images reconstructed using the smoothing prior and the median root prior had slightly lower contrast than the standard reconstruction with collimator correction, whereas the anatomical prior produced images with good resolution and contrast.
机译:单光子发射计算机断层扫描(SPECT)的较差分辨率降低了其在临床实践中的使用。已显示准直仪校正可改善重建的分辨率,但校正会产生振铃伪影,从而降低图像质量。本文研究贝叶斯重构方法是否可以减少这些伪像。我们已经应用并测试了三种贝叶斯重建方法:平滑先验,中值根先验和解剖学先验。为了证明这些方法的有效性,我们在幻像和患者研究中比较了它们的物理和视觉性能。所有三种贝叶斯重建方法都减少了准直仪校正伪影。与使用准直仪校正的标准重建相比,使用平滑先验和中位先验重建的图像的对比度稍低,而解剖学先验产生的图像具有良好的分辨率和对比度。