...
首页> 外文期刊>Earth, planets and space: EPS >Development of airglow temperature photometers with cooled-CCD detectors
【24h】

Development of airglow temperature photometers with cooled-CCD detectors

机译:带有冷却CCD检测器的气辉温度光度计的开发

获取原文
           

摘要

We have developed three airglow temperature photometers with cooled-CCD detectors. The photometers measure rotational temperatures using the airglow emissions of OH and O2 near the mesopause region (altitude: 80-100 km). The photometers also measure six other airglow and auroral lines at wavelengths of 557.7, 630.0, 777.4, 589.3, 427.8, and 486.1 nm. The CCD detectors are used to distinguish the emission lines in these airglow bands, similarly to those used by the Spectral Airglow Temperature Imagers (SATI). In this paper, we describe the configuration of the photometers, their calibration, the data processing to extract rotational temperatures and emission intensities from the measured airglow spectra, as well as the initial deployment at Platteville, Colorado (40.2°N, 255.3°E), when their observations were compared with the concurrent and nearly collocated observations by a sodium lidar. We obtain a good correlation and some systematic difference of temperatures from the photometers and the lidar, and discuss possible causes of the temperature difference.
机译:我们已经开发了三种带有冷却CCD检测器的气辉温度光度计。光度计使用中空区(海拔:80-100 km)附近的OH和O 2 的辉光发射来测量旋转温度。光度计还测量其他六条气辉和极光线,其波长分别为557.7、630.0、777.4、589.3、427.8和486.1 nm。 CCD检测器用于区分这些气辉波段中的发射线,类似于光谱气辉温度成像仪(SATI)所使用的那些。在本文中,我们描述了光度计的配置,它们的校准,从测得的气辉光谱中提取旋转温度和发射强度的数据处理,以及在科罗拉多州普拉特维尔(40.2°N,255.3°E)的初始部署,当他们的观测值与钠激光雷达的同时观测值和并列观测值进行比较时。我们从光度计和激光雷达获得了良好的相关性和一些系统的温度差异,并讨论了造成温度差异的可能原因。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号