首页> 外文期刊>Journal of Diabetes Science and Technology >Manufacturing Process Changes and Reduced Skin Irritations of an Adhesive Patch Used for Continuous Glucose Monitoring Devices
【24h】

Manufacturing Process Changes and Reduced Skin Irritations of an Adhesive Patch Used for Continuous Glucose Monitoring Devices

机译:用于连续血糖监测设备的胶贴的制造工艺变更和减少的皮肤刺激

获取原文
           

摘要

Adhesives used for continuous glucose monitoring (CGM) devices can cause skin irritations, which sometimes lead to abandonment of the therapy. A previous sensor manufacturing process involved two separate adhesives—one applied to the skin-facing surface of the fabric patch, and a second, ethyl cyanoacrylate-based adhesive, which secured the plastic transmitter housing to the superficial side of the patch. Our current process for attaching the transmitter housing to the fabric patch uses heatstaking, wherein the housing is heated and pressed against the patch with a specialized assembly apparatus. Heatstaking simplifies the sensor assembly process and obviates the need for the second adhesive, which may lead to lower risk of skin irritation(s) in some patients.
机译:用于连续葡萄糖监测(CGM)设备的胶粘剂会引起皮肤刺激,有时会导致放弃治疗。以前的传感器制造过程涉及两种单独的粘合剂-一种涂在织物贴片的面向皮肤的表面上,另一种是氰基丙烯酸乙酯基粘合剂,用于将塑料变送器外壳固定到贴片的表面。我们当前将发射器外壳连接到织物补片上的过程采用热熔法,其中将外壳加热并用专用的组装设备压在补片上。散热简化了传感器的组装过程,并且消除了对第二种粘合剂的需求,这可能导致某些患者的皮肤刺激风险降低。