In this work we report the design, implementation and results of an alternative monitoring system for a linear accelerator (LINAC) used in medical therapy. The system proposed consist in aslab of scintillator plastic with awavelength shifter fiberoptically coupled to collect the light generated,and a Spectrophotometer Ocean Optics USB4000 as analyzer. The control was made with two computers, one into the therapy room and another, using a VNC (Virtual Network Computer) and Ethernet wire, outside of the room in order to avoid radiation exposure. The LINAC dose range covered was 1, 2, 3, 4, 5, 10, 20, 30 and 40 Monitor Units (MU) with 6 and 18 MeV energy photons. The spectrum obtained was compared with the measures of the LINAC ionization camera used to calibrate it. The results obtained allowus to propose this device as an alternative method to monitor the LINAC performance.
机译:在这项工作中,我们报告了用于医学治疗的线性加速器(LINAC)的替代监视系统的设计,实施和结果。所提出的系统包括闪烁体塑料的实验室,该闪烁体塑料具有光纤耦合的波长移位器以收集产生的光,并使用分光光度计Ocean Optics USB4000作为分析仪。控制是由两台计算机进行的,其中一台进入治疗室,另一台通过VNC(虚拟网络计算机)和以太网线进入房间外,以避免辐射暴露。涵盖的LINAC剂量范围是具有6和18 MeV能量光子的1、2、3、4、5、10、20、30和40个监视单元(MU)。将获得的光谱与用于校准它的LINAC电离相机的测量值进行比较。获得的结果令人鼓舞,建议将该设备推荐为监视LINAC性能的替代方法。