首页> 外文期刊>Journal of the Brazilian Chemical Society >Analytical performance of an inductively coupled plasma optical emission spectrometry with Dual View configuration
【24h】

Analytical performance of an inductively coupled plasma optical emission spectrometry with Dual View configuration

机译:具有Dual View配置的电感耦合等离子体发射光谱的分析性能

获取原文
           

摘要

The analytical performance of axially and radially viewed inductively coupled plasma optical emission spectrometers (AX-ICP OES and RD-ICP OES, respectively) were evaluated in terms of the Mg II/Mg I ratio, matrix effects, and detection limits (LOD). All solutions were made in either 1% v v-1 nitric acid or in a 10% v v-1 water-soluble tertiary amine solution (CFA-C) adjusted to pH 8. Using a nebulization gas flow-rate of 0.5 L min-1 and applied power of 1.3 kW, Mg II/Mg I ratio was at least 15 in any medium for both configurations. Under robust conditions, plasma processes were practically not affected by the presence of 0.1% m v-1 of Ca, K, or Na in both configurations. However, interference effects were observed with non-robust conditions, mainly with AX-ICP OES in both media. The best detection limits were obtained using AX-ICP OES under robust conditions, with solutions prepared in the amine medium.
机译:根据Mg II / Mg I之比,基质效应和检测限(LOD)评估了轴向和径向观察的电感耦合等离子体发射光谱仪(分别为AX-ICP OES和RD-ICP OES)的分析性能。所有溶液均在1%v v-1硝酸或10%v v-1水溶性叔胺溶液(CFA-C)中调节至pH8。使用0.5 L min的雾化气体流速-1和施加功率为1.3 kW的Mg II / Mg I比率在两种配置中的任何介质中均至少为15。在稳定的条件下,两种配置中实际上都不受0.1%m v-1的Ca,K或Na的存在影响等离子体工艺的影响。但是,在非鲁棒条件下(主要是在两种介质中使用AX-ICP OES)都观察到了干扰效应。使用AX-ICP OES在稳定条件下,在胺介质中制备溶液,可获得最佳检测限。