首页> 外文期刊>Pediatrics: Official Publication of the American Academy of Pediatrics >Relative Efficacy of an Incubator and an Open Warmer in Producing Thermoneutrality for the Small Premature Infant
【24h】

Relative Efficacy of an Incubator and an Open Warmer in Producing Thermoneutrality for the Small Premature Infant

机译:保育箱和开放式保温柜在产生早产儿热中性的相对功效

获取原文
           

摘要

To determine which warming system more closely approximates a neutral thermal environment, the oxygen consumption of 16 premature babies, weighing less than 1,500 gm each, was measured in a convectively heated incubator and under an open radiant warmer. Both systems were controlled to maintain a skin temperature of 36 C. The oxygen consumption of the infants was significantly higher under the radiant warmer. This result is in agreement with published data for this group of premature infants.
机译:为了确定哪种加热系统更接近中性热环境,在对流加热的培养箱中和开放辐射加热器下测量了16个早产婴儿的氧气消耗,每个婴儿的体重不足1,500 gm。两种系统均被控制为维持36°C的皮肤温度。在辐射变暖的情况下,婴儿的氧气消耗量明显更高。该结果与该组早产婴儿的公开数据相符。