【24h】

Early metallosis-related failure after total knee replacement

机译:全膝关节置换术后早期与金属化有关的衰竭

获取原文
           

摘要

Metallosis is a rare cause of failure after total knee replacement and has only previously been reported when there has been abnormal metal-on-metal contact. We describe 14 patients (15 knees) whose total knee replacement required revision for a new type of early failure caused by extensive metallosis. A modification of a cementless rotating platform implant, which had previously had excellent long-term survival, had been used in each case. The change was in the form of a new porous-beaded surface on the femoral component to induce cementless fixation, which had been used successfully in the fixation of acetabular and tibial components. This modification appeared to have resulted in metallosis due to abrasive two-body wear. The component has subsequently been recalled and is no longer in use. The presentation, investigation, and findings at revision are described and a possible aetiology and its implications are discussed.
机译:金属病是全膝关节置换术后罕见的失败原因,并且以前仅在金属与金属之间异常接触时才报道。我们描述了14例患者(15膝),他们的全膝关节置换术需要对因广泛的金属病引起的新型早期衰竭进行修订。在每种情况下,都使用过一种非骨水泥旋转平台植入物的改良,该植入物以前具有出色的长期存活率。改变的形式是在股骨组件上形成了一个新的多孔珠状表面,以诱导非骨水泥固定,这种固定方法已成功用于髋臼和胫骨组件的固定。由于两个身体的磨损,这种修饰似乎导致了金属化。该组件随后被召回,不再使用。描述,研究和修订发现,并讨论可能的病因及其含义。