首页> 外文期刊>Journal of Clinical Microbiology >Impact of DNA Polymerases and Their Buffer Systems on Quantitative Real-Time PCR
【24h】

Impact of DNA Polymerases and Their Buffer Systems on Quantitative Real-Time PCR

机译:DNA聚合酶及其缓冲液系统对实时定量PCR的影响

获取原文
           

摘要

An investigation of the influence of five DNA polymerase-buffer systems on real-time PCR showed that the choice of both DNA polymerase and the buffer system affected the amplification efficiency as well as the detection window. The analytical repeatability of the data for different systems changed clearly, leading us to conclude that basing quantitative measurements on single-data-set standard curves can lead to significant errors.
机译:对五个DNA聚合酶-缓冲液系统对实时PCR的影响进行的研究表明,DNA聚合酶和缓冲液系统的选择都会影响扩增效率以及检测窗口。对于不同系统,数据的分析可重复性发生了明显变化,使我们得出结论,基于单一数据集标准曲线进行定量测量可能会导致重大错误。