Background Kit lot change in clinical biochemistry labs leads to variations in patient results. This study planned to identify variations during 60 reagent lot changes in our laboratory during the period from June 2018 to May 2019. Methods A statistical analysis was performed to identify the difference between patient samples results variations and QC results. The long term drift was analyzed using a regression test. Results There was a significant difference between the patient and QC results in 16.7% of reagent lot changes. Moreover, the extent of variation in QC results was 3.3%. No long-term drift was seen in three analytes which were studied using regression analysis. Conclusions Our results showed that, during reagent kit lot change, along with QC material, the patient samples should also be run in order to identify the variation. However, this practice is presently ignored by most of the laboratories. There was no accumulated effect in our laboratory due to reagent kit lot change.
机译:背景包批次改变临床生物化学实验室导致患者结果的变化。本研究计划在2018年6月至2019年5月期间,在我们实验室期间识别我们实验室的60个试剂批次变化的变化。方法进行统计分析以确定患者样品结果变异和QC结果之间的差异。使用回归测试分析长期漂移。结果患者和QC之间存在显着差异,导致16.7%的试剂批次变化。此外,QC结果的变化程度为3.3%。在使用回归分析中研究的三种分析物中没有看到长期漂移。结论我们的结果表明,在试剂套件批次期间,以及QC材料,也应运行患者样品以识别变异。但是,大多数实验室都忽略了这种做法。由于试剂套件改变,我们的实验室没有累积效应。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号