首页> 外文期刊>Brazilian Journal of Anesthesiology >Oxygen delivery failure due to improper installation of carbon dioxide absorbent canister: a case report
【24h】

Oxygen delivery failure due to improper installation of carbon dioxide absorbent canister: a case report

机译:由于安装二氧化碳吸收罐的不当原因,氧气输送失效:案例报告

获取原文
           

摘要

The CLIC system in the Dr?ger Apollo anesthesia workstation allows a successful pre-use machine checkout without the presence of a carbon dioxide absorbent canister. It also allows the canister to be changed without interrupting controlled ventilation. However, this canister can be easily installed improperly with the CLIC adapter. We report a case in which a patient could not be ventilated by mask after the induction of general anesthesia, resulting in oxygen desaturation before successful ventilation was achieved with a bag valve mask. This case illustrates the importance of a leak test after components of the breathing circuit are changed.
机译:DR的CLIC系统?GER Apollo麻醉工作站允许成功的预使用机器结账而不存在二氧化碳吸收罐。 它还允许罐改变而不中断受控通风。 但是,本罐可以轻松地使用Clic适配器安装不正确。 我们报告了一种情况,其中患者在诱导全身麻醉后不能通过掩模通风,导致在用袋阀掩模实现成功通风之前的氧气去饱和。 这种情况说明了在呼吸电路的组件改变之后泄漏测试的重要性。